Before Log. Pressing Log button zeros the sensor.
Start Pressed. Pressing Start button zeros the sensor.
36
Chapter 5 Settings
Manual. When Manual is selected, the relay remains in its present state. The user can
close (turn on) or open (turn off) the relay at will.
Start. Relay is closed (turned on) as soon as the process starts.
Predeposit. Relay is closed (turned on) at the beginning of the predeposit phase during
the predeposition process.
Auto Deposition. Relay is closed (turned on) just before the PID starts, activating as
soon as Dwell Time (preset) is initiated. Once Dwell Time concludes, Auto Deposition
begins. This process is designed to prevent the Automatic Deposition from engaging
prematurely, providing a window of time between the shutter opening and the material
reaching the substrate. This selection is specically designed for shutters.
Mirage™ (Relay 2 Only). When the heater is active, Eon™ uses Relay 2 to control
Mirage™. In order to activate the heater, the Relay 2 Mirage option must be selected.
While the heater is active, Relay 2 settings cannot be changed (unless the user is in
Manual mode [for more about Manual mode, see Chapter 6]).
Sensor Zeroing
With the Sensor Zeroing panel, the user can select when to zero Sensor 1 or Sensor 2.
Settings for each sensor are identical.
Sensors Tab