CAREFULLY READ AND UNDERSTAND INSTRUCTIONS BEFORE USING CHAIR
Warning!
The battery room is supplied with ventilation that ensures correct and necessary airing of the battery
room. This airing must not be blocked or covered as a positive pressure may occur with risk of
explosion.
If the product has been exposed to mechanical overload (lost on the floor, collision/squeezing in the
application or a powerful stroke) the product must be sent to an authorised workshop for control of
the hermetic separation between the battery and electronics rooms.
CLEANING & DISINFECTION of ELECTRICAL COMPONENTS
The products can be cleaned as follows according to their IP protection, which is stated on the product
label:
Wash with a brush and water, but not water under pressure.
Wash with a brush and water. The water can be under pressure, but the system
must not be hosed down directly with a high pressure cleaner. Max. 20
°
C.
Clean by the use of water tunnels
X can have a value of between 1 and 6.
* The following products can be marked with IPX6, but they are still allowed for wash tunnels: ACP,
ACM (mini), ACM (new), BA18, CB12/14/18, DJB, FS2, HB40, LA28, LA32, LA34 stand, LA34 comp.
The materials are resistant to the majority of cleaners and disinfectants used in the hospital and
nursing home sector. Cleaning with a steam cleaner, for example, is not permitted, as it will not be
possible to maintain a minimum distance of 300 mm from the electrical parts.
IP 66W products can be washed in wash tunnels that wash according to the “Machine
Decontamination” provided that the following guidelines are complied with:
The cleaning and disinfection cycle in the wash tunnel must not last longer than 10 minutes.
The nozzle pressure in the wash tunnel must not exceed 10 bar.
The distance between the nozzle and the electrical parts must be at least 300 mm.
The handset and motor plug must be pushed right in.
The water temperature must not exceed 85°C.
Cooling with cold water is not permitted.
To avoid degreasing of the piston rod, the actuator should be retracted to minimum stroke
before washing.
The following directions regarding cleaners and disinfectants must be complied with:
• They must not be highly alkaline or acidic (pH value 6-8).
• They must not contain caustic agents.
• Their contents must not be able to change the structure of the surface or adhesion of the
plastic.
7. CLEANING & MAINTENANCE