Manual Electrophoresis Power Supply, rev. 32/2018
9
Manual programming
Manual programming permits to set voltage, current, power and time limits for a simple routine
electrophoresis run. Parameters can be changed temporarily without interrupting the run.
This menu is automatically shown in the EV0220 after start-up.
Verify if the instrument is switched off.
Connect the electrophoresis tank to the instrument.
Switch the instrument on.
Select MANUAL and press :
(not EV0220)
Enter the desired voltage limit
and press :
Enter the desired current limit
and press :
Enter the desired power limit
and press :
Enter the desired timer unit (h or kVh)
and press :
h
The timer will count in hours and
minutes.
kVh
The timer will count in kilovolthours.
Enter the desired timer value
and press :
h
The value format is hh:mm.
kVh
The format has a value with 2 decimal
places..
Set to 00:00 when required to
continue until / is pressed or the
device is switched off.
Press / to start the experiment.
At any time, you can return to the main menu by holding pressed.
At any time, you can avoid programming unnecessary parameters. To do so, simply press /
instead of after entering the last relevant value and the instrument will immediately start the
experiment keeping the other parameters to their previous values.
To ensure that the desired parameter is kept constant, we suggest to preset all other parameters to
its maximum.
The behaviour of the power supply during a procedure is described in the chapter Running Condition.
PROCEDURE?
MANUAL
METHOD
SETUP
MANUAL
VOLTAGE
300
V
MANUAL
CURRENT
500
mA
MANUAL
POWER
150
W
MANUAL
TIMER UNIT
h
MANUAL
TIMER
00:30
h