Environmental condition of use
COSMED units should not be operated near explosive substances.
Equipment should not be installed near electrical or magnetic devices such as x-ray equipment, transformers or power lines. These devices could
create electrical interferences when performing testing procedures. COSMED devices are not AP or APG units (according to EN 60601-1) and should
never be operated in the presence of flammable anaesthetic mixtures.
COSMED equipment should be operated under normal environmental temperatures and conditions which are defined as follows [IEC 60601-1/EN
60601-1]:
Temperatures range: 10°C (50°F) and 40°C (104°F).
Relative humidity range: 30% to 90% (not condensing).
Atmospheric Pressure range: 600 mBar to 1060 mBar.
Avoid operating equipment in the presence of noxious fumes or in dusty environments.
Do not place units near heat sources.
Cardiopulmonary resuscitation equipment should be accessible in the case of an emergency.
Adequate floor space and easy access to the patient during exercise testing is necessary.
Adequate ventilation should be maintained in the room the testing is performed.