4. In order to let it dry, please shake it in air and/or connect it to the calibration syringe through the antibacterial filter and perform some
strokes.
5. After cleaning the turbine, check if the turbine propeller rotates freely even with a low speed air flow.
6. Connect the turbine to the reader.
Precautions to take when cleaning, disinfecting and drying the turbine
Do not expose the turbine to high heat or to a direct flow of water.
Do not expose the sampling tube or the connector on the end of the cable to any liquids.
Do not use alcoholic solutions to clean the turbine.
28 - System maintenance - Cleaning and disinfecting