S Series
SERVICE MANUAL

Questions and Answers

  • C
    Connie YoungAug 1, 2025
    What to do if my Covaris S Series shows no power?
    • N
      nixonallisonAug 1, 2025
      If your Covaris Laboratory Equipment isn't powering on, first ensure the line cord is properly connected. If it is, check and replace the fuse S1/S2 in the line entry module. Also, check the Green LED2 on tuning board S220 and Green D1 on the Controller Board.
  • D
    Dr. Angela NelsonAug 5, 2025
    Why is my Covaris S Series ineffective and making a loud buzzing noise?
    • J
      Jennifer HardyAug 5, 2025
      Ineffective treatment and buzzing from your Covaris Laboratory Equipment can occur if the bath temperature is too high, the bath is contaminated, not sufficiently degassed, or there is insufficient acoustic power. Verify chiller operation. Clean the bath, change the water, and degas it, checking the oxygen level with a DO meter. Run an acoustic power test.
  • M
    Mark SmithAug 10, 2025
    What to do if my Covaris S Series Laboratory Equipment displays a Water Level Error?
    • M
      morrownatalieAug 10, 2025
      A water level error on your Covaris Laboratory Equipment may be due to insufficient bath water, a contaminated level sensor, ambient light hitting the sensor, or a faulty sensor. Fill the bath to the indicator line, clean the bath and level sensor, shield the sensor from stray light, and if necessary, replace the sensor.
  • S
    Sheri HarperAug 14, 2025
    What to do if my Covaris Laboratory Equipment shows a Communication Error?
    • G
      Gregory HardyAug 14, 2025
      A communication error with your Covaris Laboratory Equipment can stem from corrupted USB drivers, incorrect tuning board switch settings, a defective USB cable, or using a USB cable that is too long. Reload the drivers, verify the correct switch settings, replace the USB cable, and ensure it does not exceed 2 meters in length.
  • T
    Trevor SweeneyAug 19, 2025
    How to fix Safety Interlock Error on Covaris S Series Laboratory Equipment?
    • C
      Chelsea SimmonsAug 19, 2025
      To resolve a Safety Interlock Error on your Covaris Laboratory Equipment, verify the bath position and ensure that the lid of the acoustic bracket is fully down.
  • T
    Taylor WoodardAug 23, 2025
    Why are there excessive degassing bubbles in my Covaris S Series?
    • A
      ashley13Aug 23, 2025
      Excessive degassing bubbles in your Covaris Laboratory Equipment may be caused by the inlet tube not being properly seated or a leak in the degas pump tubing or fitting. Make sure the inlet tube is inserted past the O-ring and check the integrity of the tubing and connections.
  • C
    Christopher OwensAug 27, 2025
    What to do if my Covaris Laboratory Equipment has no degassing bubbles?
    • J
      joseph49Aug 27, 2025
      If your Covaris Laboratory Equipment is producing no degassing bubbles, the degas inlet nozzle might be clogged, the degas pump may not be running, or the motor could be decoupled from the pump. Try cleaning the bath and degas pickup tube. Verify the water level and pump operation. Stop and restart the degassing pump.

Summary

Warnings

For Safety of Operating Personnel

Essential safety precautions for personnel operating the instrument.

To Prevent Damage to the Equipment

Guidelines to prevent damage to the Covaris instrument during operation.

Safety and EMC Compliance Information

General Instrument Safety

General safety guidelines for operating the Covaris S-Series instruments.

Introduction

Transducer

Explanation of the transducer's function and role in acoustic processing.

Degassing System

Details on the system's degas pump and its function for acoustic processing.

Maintenance and Repair

Recommended Daily Maintenance

Daily checks and procedures for maintaining the instrument's performance.

Troubleshooting

Guide to identifying and resolving common instrument problems and symptoms.

Repair Procedures

Step-by-step instructions for performing various repair procedures on the instrument.

Diagnosing and Replacing Printed Circuit Boards: S220 Series

Procedures for diagnosing and replacing PCBs specific to the S220 models.

Acoustic Power Testing Procedure

S1;S2 Series

Procedure for measuring acoustic power output on S1/S2 models.

S220 Series

Procedure for measuring acoustic power output on S220 models.

Installing SonoLab Software

S1; S2 Series: SonoLab (Version 2)

Procedure for installing SonoLab software and USB drivers on Windows XP.

S220 Series: SonoLab 7

Procedure for installing SonoLab software and USB drivers on Windows XP or 7.

System Wiring Diagram

S1, S2 Series

Wiring diagram for S1 and S2 Series instruments.

S220 Series

Wiring diagram for S220 Series instruments.

Covaris S Series Specifications

General IconGeneral
BrandCovaris
ModelS Series
CategoryLaboratory Equipment
LanguageEnglish