SERVICE MANUAL

Questions and Answers

  • M
    Michele MercadoAug 5, 2025
    What to do if my Covaris Laboratory Equipment shows no power?
    • L
      lindavaldezAug 5, 2025
      If your Covaris Laboratory Equipment isn't powering on, first ensure the line cord is properly connected. If it is, check and replace the fuse S1/S2 in the line entry module. Also, check the Green LED2 on tuning board S220 and Green D1 on the Controller Board.
  • T
    Tara HuntAug 9, 2025
    Why is my Covaris S2 ineffective and making a loud buzzing noise?
    • J
      James PattonAug 10, 2025
      Ineffective treatment and buzzing from your Covaris Laboratory Equipment can occur if the bath temperature is too high, the bath is contaminated, not sufficiently degassed, or there is insufficient acoustic power. Verify chiller operation. Clean the bath, change the water, and degas it, checking the oxygen level with a DO meter. Run an acoustic power test.
  • D
    Dr. Brandon Rowe DDSAug 14, 2025
    What to do if my Covaris S2 displays a Water Level Error?
    • M
      mckinneytravisAug 14, 2025
      A water level error on your Covaris Laboratory Equipment may be due to insufficient bath water, a contaminated level sensor, ambient light hitting the sensor, or a faulty sensor. Fill the bath to the indicator line, clean the bath and level sensor, shield the sensor from stray light, and if necessary, replace the sensor.
  • R
    Rachael HuntAug 18, 2025
    What to do if my Covaris Laboratory Equipment shows a Communication Error?
    • S
      Stacie ThorntonAug 18, 2025
      A communication error with your Covaris Laboratory Equipment can stem from corrupted USB drivers, incorrect tuning board switch settings, a defective USB cable, or using a USB cable that is too long. Reload the drivers, verify the correct switch settings, replace the USB cable, and ensure it does not exceed 2 meters in length.
  • J
    Jose NunezAug 22, 2025
    How to fix Safety Interlock Error on Covaris S2?
    • W
      William DawsonAug 23, 2025
      To resolve a Safety Interlock Error on your Covaris Laboratory Equipment, verify the bath position and ensure that the lid of the acoustic bracket is fully down.
  • G
    gabrielharrisAug 27, 2025
    Why are there excessive degassing bubbles in my Covaris S2?
    • E
      Erin ClarkAug 27, 2025
      Excessive degassing bubbles in your Covaris Laboratory Equipment may be caused by the inlet tube not being properly seated or a leak in the degas pump tubing or fitting. Make sure the inlet tube is inserted past the O-ring and check the integrity of the tubing and connections.
  • T
    Tonya StewartAug 30, 2025
    What to do if my Covaris Laboratory Equipment has no degassing bubbles?
    • A
      Angelica AdamsAug 30, 2025
      If your Covaris Laboratory Equipment is producing no degassing bubbles, the degas inlet nozzle might be clogged, the degas pump may not be running, or the motor could be decoupled from the pump. Try cleaning the bath and degas pickup tube. Verify the water level and pump operation. Stop and restart the degassing pump.

Summary

Universal Precautions

Warnings

For Safety of Operating Personnel

Instructions for grounding, power cord, and safety system operation.

To Prevent Damage to the Equipment

Guidelines for operating conditions, water bath maintenance, cleaning, and software use.

Maintenance and Repair

Recommended Daily Maintenance

Daily tasks for filling and emptying the water bath.

Troubleshooting

Table summarizing common symptoms and their corrective actions.

Diagnosis

Detailed diagnostic procedures for common issues.

Repair Procedures

Step-by-step guides for various repair tasks.

Pump Replacement

Detailed steps for replacing the degas pump assembly.

Diagnosing and Replacing Printed Circuit Boards: S1 & S2

RF Generator board

Details the RF Generator board, its functions, and failure modes.

Diagnosing and Replacing Printed Circuit Boards: S220 Series

RF Power Board

Details the S220 RF Power Board and its unique calibration data.

Power Supplies

SonoLab Service Menus; Access

Service Level Login

Steps for logging into SonoLab with service credentials.

Diagnosis

Water Bath and Degassing Issues

Diagnosing problems related to water bath conditions and degassing.

Contamination

Bath Temperature

Water Level Errors

Acoustic Power Issues

Interlock Errors

Communication Errors

Repair Procedures

General: Removing the Covers (continued)

Pump Replacement

Pump Replacement (continued)

Replacing the Water Sensor

Replacing the Water Sensor (continued)

Replacing the Transducer Assembly

Replacing the Transducer Assembly (continued)

RF Generator board (continued)

RF generator board installation and switch configuration

S-Tuning board

RF Power Board

Main Circuit Subsections (continued)

RF Power Board Replacement

Controller Board

Controller Board Replacement (continued)

Match Network PCBA and Transducer

Replacement of 220 Series Match Network PCBA and Transducer (continued)

S1; S2 Series (continued)

S220 Series

24 Volt Supply

48 Volt Supply

48 Volt Supply (continued)

S220 Power Supply Replacement

Updating Transducer Information after Repair

Updating Transducer Information after Repair (continued)

Updating Unit Information after Repair (continued)

Acoustic Power Testing Procedure

S1; S2 Series

Steps for creating a treatment and measuring acoustic power.

Installing SonoLab Software (Subject to Change)

S220 Series: SonoLab 7

Covaris S2 Series Specifications

General IconGeneral
BrandCovaris
ModelS2 Series
CategoryLaboratory Equipment
LanguageEnglish