EasyManua.ls Logo

Covidien Shiley - Tracheostomy Tube Insertion

Covidien Shiley
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
2. With the Shiley Disposable Cannula Cued models (DCT, DFEN) the cu and
ination system should be tested for leakage before inserting the tube. This test can be
performed as follows: Inate the cu with the volume of air indicated in Tables 1 and
2. Then either observe for deation over several minutes or immerse the tube in sterile
saline and observe for air leakage. Deate the cu prior to insertion.
Insertion
CAUTION:
To ease insertion and to guard against cu perforation from sharp edges of
cartilage, the cu should be tapered back. This can be accomplished by rst
inating the cu. Then gently move the cu away from the distal tip of the
outer cannula towards the swivel neck plate as the residual air is removed by
deation. Do not use any sharp instruments such as forceps or hemostats that
would damage the cu when tapering it.
3. Remove the disposable inner cannula and insert the obturator into the outer
cannula. The obturator should be fully seated before the tube is inserted into the
patient. A thin lm of water soluble lubricant can be applied to the outer cannula, cu,
and protruding portion of the obturator to facilitate insertion.
4. After the tracheostomy procedure is performed, insert the tube into the patient’s
trachea. After the tube is properly in place, remove the obturator immediately.
CAUTIONS:
When using a water soluble lubricant verify that the lubricant does not enter
and occlude the tube lumen which may restrict ventilation.
The obturator should be cleaned and then stored in an accessible location near
the patient should it be needed for an unscheduled re-intubation.
WARNINGS:
THE SHILEY DISPOSABLE INNER CANNULA PROVIDED IN THIS PACKAGE CAN
ONLY BE REPLACED BY A SHILEY DISPOSABLE INNER CANNULA (DIC) OF THE
SAME SIZE NUMBER.
A DISPOSABLE INNER CANNULA (DIC) SHOULD BE WITH THE PATIENT AT ALL
TIMES. IN THE EVENT OF RESPIRATORY DISTRESS, THE DISPOSABLE INNER
CANNULA (DIC) MUST BE INSERTED TO MECHANICALLY VENTILATE THE PATIENT.
DURING AND AFTER ATTACHMENT OF RESPIRATORY OR ANESTHESIA
TUBING AND/OR CONNECTORS TO THE DISPOSABLE INNER CANNULA, AVOID
APPLICATION OF EXCESSIVE ROTATIONAL, LINEAR, OR ROCKING FORCES ON
THE TUBING AND/OR CONNECTORS TO PREVENT ACCIDENTAL DISCONNECTION
OF THE DISPOSABLE INNER CANNULA, OR DAMAGE TO THE TRACHEOSTOMY
TUBE.
5. Insert the disposable inner cannula with 15 mm connector and lock into position.
To lock the disposable inner cannula in place, push connector until both snap-locks

Related product manuals