11
Tetra Event logging
Peak test
When performing a peak test, such as a vertical entry check, previous
readings can be cleared on entry to the peak display option.
Zero
The Tetra instrument can perform an autozero by selecting the Zero
function from the menu. When zero is completed, the instrument will
return to normal operation.
2.4 Logging
Tetra incorporates event logging which can be accessed using the IR communi-
cations link with Portables PC software. See section VIII.
Tetra records the time and date for a number of operating and diagnostic
events including:
Switch on and switch off
Level 1, Level 2 and Time Weighted Average Alarms, alarm on,
alarm off and the peak level during the alarm
Zero, calibration and gas test with success or failure
Catalytic bead saver on and off
The battery condition is logged every 15 minutes while the
instrument is operating, and certain configuration changes are
also logged.