EasyManua.ls Logo

Crown D-150 - Checkout Procedure - J KM2 Input; with Load

Crown D-150
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
%
SYMPTOM
Output
unstable
when
adjusting
level
pot
PROBABLE
CAUSE
AND
REMEDY
positive
check
the
+
side.
If
offset
is
negative
check
the
+
side.
Defective
bias
diodes
(02-D5).
Replace.
:
NOTE;
As
a
general
procedure,
use
a
voltmeter
to
compare
voltage
readings
with
those
shown
on
the
unit
schematic.
Begin
at
the
input,
move
stage
by
stage
through:
the
circuitry,
and
note
any
Jarge.deviations.
from
the
quiescent
vollage
levels
shown.
Determine
whether
specific
components
are
defective
or
simply
responding
to
a
problem
elsewhere.
Check
.componerits-on
Main.
PC
Board
by
alternately
heating
and
cooling
(use
small
sokier
ivon
and
circuit
cooler
spray)
while
observing
the
oscillation
on
the
oscilloscope,
Concen-
trate
on
transistors
and
diodes.
Significant
changes
in
the
oscilla-
tion
may
pin-point
a
defective
component.
Check
diodes
and
resistors
by
paralleling
witha
capacitor.
This
may
help
isolate
the
source
of
an
unwanted
oscillation.
CAUTION!
In
some
cases
this
technique
may
cause
oscillation
so
use
with
care.
Check
capacitors
by
paralleling
ther
with
a
known
good
capac-
itor
of
the
same
value.
1
oscilla-
tion
is
eliminated,
an
open
capacitor
may
be
indicated.
Dirty
internal
contacts
on
level
pot.
Temporary
correction
of
this
problem
may
sometimes
be
had
by
turning
the
pot
back
and
forth
repeatedly
to
allow
the
contacts
to
clean
them-
selves.
If
this
is
unsuccess-
ful,
replace
the
pot.
Sarre
me-esewrweneienienntinihasiteerensineshonimpstnanp
neta
Sirniiielies
25
5.7.5
Checkout
Procedure
1
KHz
Input,
With
Load
Most
of
the
common
service
problems
should
have
manifested
themselves
previously
and
the
remaining
test
procedures
shoukt
be
fairly
routine.
if
troubles
do
occur,
a
likely
source
is
the
output
Stages,
since
this
is
the
first
test
where these
stages
are
subjected
toa
load.
1.
Connect
an
8-ohm
load
to
the
output.
(Refer
to
paragraph
§.6.2).
2.
Monitor
the
output
with
an
oscilloscope
and
an
accurate
AC
~
voltmeter.
3.
Connect
a
sine
wave
oscillator
set
at-1
KHz
(+10dB
out)
to
the
amplifier
input.
4.
Turn
up
the
level
control
of
the
channel
under
test:
The
output
must
clip
at
over
98
watts
(28
volts).
Typically,
the
clip
point
is
105
watts
or
above,
with
the
ac
line
voltage
at
120
volts.
The
waveform
should
be
clean,
and
should
clip
on
and
ar
matrically.with
ne.
ringing.
;
5.
Change
load
to
4
onms.
Clip
point
should
occur
at
approx:
imately
170
watts
(26
volts):
:
5.7.6
Checkout
Procedure
Protection
|
Circuit
Tests
,
The
protection
circuit
test
is
a
clipping
test
designed
to
verity
the
operation
of
the
limiting
circuits.
1.
Set
the
amplifier
output
at
approximately
28V
across
8
ohms.
2.
Switch
the
foad
to
2
ohms.
At
this
load
impedance
the
output
will
cause
enough
power
supply
sag
to
clip
the
output
but
may
not
cause
the
sharp'clip
indicative
of
limiter
circuit
action.
Slowing
the
oscilloscope
trace
to
look
for
power
supply
ripple
at
the
clip
level
will
indicate
whether
the
power
supply
or
limiter
circuits
are
causing
the
clip.
3.
in
order
to
insure
that
the
limiter
circuits
are
operating,
it
may
be
necessary
to
driva
a
1-ohmn
load.
If
the
oscilloscope
trace
still
shows
only
power
supply
clipping,
the
associated
limiters)
is
not
turning
on
and
the
limiter
circuitry
is
defective.
PROBABLE
CAUSE
AND
REMEDY
SYMPTOM
+
signal
does
not
clip
Check
components
in
+
limiter
on
2-ohm
load
(Q6
circuit).
4
-
signal
does
not
clip
Check
components
in
-
limiter.
on
2-ohm
load
(Q7)
circuit.
Either
+
or
-
signal
Check
limiter
circuits.
Refer
to
clips
at
wrong
level
Figure.
5-2
for
clipping
levels.
‘on
2-ohm
load.
neater
nininnnenine
PATA
EREB
EN
BAIRIRAL.

Other manuals for Crown D-150

Related product manuals