EasyManua.ls Logo

CRYOMAGNETICS 4G - Page 47

CRYOMAGNETICS 4G
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cryomagnetics Model 4G Operating Instruction Manual 40
Table 2: Computer Interface Command Summary
Command
Available
Description
CHAN
Remote
Select module for subsequent commands (Dual Supply only)
CHAN?
Always
Query module selection (Dual Supply only)
ERROR
Remote
Set error response mode for RS-232 interface
ERROR?
Always
Query error response mode
IMAG
Remote
Set magnet persistent current
IMAG?
Always
Query magnet current
IOUT?
Always
Query power supply output current
LLIM
Remote
Set low current sweep limit
LLIM?
Always
Query current sweep limit
LOCAL
Always
Return control to front panel (RS-232 Only)
MODE?
Always
Query selected operating mode
NAME
Remote
Set magnet coil name
NAME?
Always
Query magnet coil name
PSHTR
Remote
Control persistent switch heater
PSHTR?
Always
Query persistent switch heater state
QRESET
Remote
Reset quench condition
RANGE
Remote
Set range limit for sweep rate boundary
RANGE?
Always
Query range limit for sweep rate boundary
RATE
Remote
Set sweep rate for selected sweep range
RATE?
Always
Query sweep rate for selected sweep range
REMOTE
Operate
Select remote operation (RS-232 Only)
RWLOCK
Operate
Select remote operation with front panel lock (RS-232 Only)
SHIM
Remote
Select shim to be queried or changed (shim option only)
SHIM?
Always
Query shim selection (shim option only)
SLIM?
Always
Query current limit for selected shim (shim option only)
SLIM
Remote
Set current limit for selected shim (shim option only)
SWEEP
Remote
Start output current sweep
SWEEP?
Always
Query sweep mode
ULIM
Remote
Set current sweep upper limit
ULIM?
Always
Query current sweep upper limit
UNITS
Remote
Select units
UNITS?
Always
Query selected units
VLIM
Remote
Set voltage limit