MNL-00008 [5] Proprietary and Confidential Page 51 of 55
Radiated, radio-frequency,
electromagnetic field
immunity test following IEC
61000-4-3
80 MHz to 2.7GHz @ 10.0
V/m
80 MHz to 2.7GHz @ 10.0
V/m
Electrical fast transient/burst
immunity test following IEC
61000-4-4
Surge immunity test
following IEC 61000-4-5
+/- 0.5 kV, +/- 1.0 kV, +/-
2kV
+/- 0.5 kV, +/- 1.0 kV, +/-
2kV
Conducted Immunity test
following IEC 61000-4-6
0.15 MHz to 80 MHz @ 3.0
Vrms
6.765 MHz to 6.795 MHz
@6.0 Vrms
13.553 MHz to 13.567 MHz
@ 6.0 Vrms
26.057 MHz to 27.283 MHz
@ 6.0 Vrms
40.66 MHz to 40.70 MHz @
6.0 Vrms
0.15 MHz to 80 MHz @ 3.0
Vrms
6.765 MHz to 6.795 MHz
@6.0 Vrms
13.553 MHz to 13.567 MHz
@ 6.0 Vrms
26.057 MHz to 27.283 MHz
@ 6.0 Vrms
40.66 MHz to 40.70 MHz @
6.0 Vrms
Power frequency magnetic
field immunity test
following IEC 61000-4-8
Voltage dips and
interruptions immunity test
following IEC 61000-4-11
30% Reduction for 500 mS
at 0 degrees,
100% Interruption for 10 mS
at 0, 45, 90, 135, 180, 225,
270,315 degrees,
100% Interruption at 5000
mS at 0 degrees
30% Reduction for 500 mS
at 0 degrees,
100% Interruption for 10 mS
at 0, 45, 90, 135, 180, 225,
270,315 degrees,
100% Interruption at 5000
mS at 0 degrees