2) Code 21
(Low pressure error)
- If pressures lower than 30
%
of the set value are detected more than 3 times during
operation, it will indicate Error code 21.
*Checking point
Check the connection of the controller to the hoses
(tubing).
Check that each cuff is wrapped properly on the
patient’s leg.
Check whether there is any leakage from a connected
cuff.
Separate the cuff and hose, check air is leaving from the air socket of the controller.
3) Code 22
(higher pressure error)
- If pressures higher than 30
%
of the set value are detected more than 3 times during
operation, it will indicate Error code 22.
*Checking points
Check whether the hose is bent.
Check if there is something blocking the hose.
Check whether air socket is blocked.
4) Code 31
(Ventilation error)
If air is not ventilated (deflated) under 20mmHg during INTERVAL TIME, Error code 31
will be indicated.
*Checking points
Check the inner solenoid valve (Exchange the valve)
5) Code 32
(Pump error)
- During the initial self detection test, if the pump is not operating normally, Error code
32 will be displayed.
*Checking points
Check pump operation after opening the cover of the controller
Check the wire connection in the controller.