4.4. Troubleshooting
Problem
Possible cause(s)
Non-zero reading
Possible presence of gas in the detector area.
Non-zero reading when no gas is present
Detector needs calibration.
Reading too high or too low
Zero reading in presence of gas
Wrong or defective electrical connection.
Sintered or sensor blocked with dust.
Sintered or sensor contaminated. Replace sin-
tered AND sensor.
4.5. Spare parts and accessories
Part or accessory
Part number
Cable gland - Ex “d” M20 (6.1 - 11.7 mm)
PRE00000032
Ferrite
FER00000001
Junction box - Ex “d”
BOI00000188
Key for locking screw (1.5 mm hex key)
OUT00000115
O-ring 17 x 2 mm
MEC00000012
Screw - Locking screw of the junction box cover
VISVIS00067
Screw - M4 x 6 mm
VISVIS00042
Screw for locking the body of sensing head
VISVIS00066
Sensing head - FPH01
*
Sensor gasket
MEC00000018
Sintered metal filter
MEC00000010
Spanner for base of sensing head (30 mm)
OUT00000114
Spanner for top of sensing head
OUT00000113
Stopper plug - Ex d M20
PRE00000033
Terminal block - Ex e 2 position
BOR00000087
Transmitter board - 4-20 mA
BASTRX00017