DLR-TL001 Functional Description
6 DLR-TL001
ETA alarm (see ETA_ALRM bit in the STATUS
register description: exceeding an extimated
time).
Bin counters alarm (see BIN_ALMR bit in the
STATUS register description: at least one histo-
gram went above its preset THR - threshold).
Mean Kinetic alarm (see MKT_ALRM bit in the
STATUS register description: temperature over a
calculated THR).
Shelf life alarm (see SHL_ALRM bit in the STA-
TUS register description: residual life time not
sufficient).
If the battery has not enough charge to continue the
logging activity, the red or green LED will blink for 2
seconds (the green one in case of no alarms, the red
one in case of some alarms).
Table 1. Alarm Indicator
If the BE bit is 1 (Start/stop function active), in addition to
the behaviour described above, the logging activity start
can be performed as follows:
•Hold down the button.
Wait until the green LED lights and the red LED
start blinking.
Release and press again the button within 5
seconds to confirm the command.
As the command is accepted, the green LED will
stay lit for 15 seconds.
The acquisition is started.
If the SD bit is 0, the logging activity stop is also allowed as
follows:
•Hold down the button.
Wait until the red LED lights and the green LED
start blinking.
NO ALARM ALARM
Temperature Status
Green Red
Battery Status
No flashing Flashing

Other manuals for Datalogic DLR-TL001