Introduction
4
DLR-TL001
Figure 1- Manual Interface
The manual interface behaves as follows:
Pressing the button quickly (less than 2
second) activates the check status
function:
• If the battery level is still sufficient for the logging activity and no
a
larms are active the green LED will stay lit for 2 seconds.
• If the battery level is still enough for the logging activity but there are
activ
e alarms the red LED will stay lit for 2 seconds. The alarms causing
the red LED to ligth are the following:
- Memory full (see MEMFULL bit in STATUS register description : see
PRG's detailed info…).
- ETA alarm (see ETA_ALRM bit in the STATUS register description :
exceeding an estimated time… ).
- Bin counters alarm (see BIN_ALMR bit i
n STATUS register descript :
at least one histogram went above its preset THR - threshold…).
- Mean Kinetic alarm (see MKT_ALRM bi
t in the STATUS register
description : temperature over a calculated THR).
- Shelf life alarm (see SHL_ALRM bit in the
STATUS register descrip-
tion: residual life time not sufficient…).
• If the battery has not enough charge to continue the logging activity the
red or green LED will blink for 2 seconds (the green one in the case of
no alarms, red in case of some alarms).