DLR-TL001 Functional Description
Product Reference Guide
5
Table 1 - Alarm
NO ALARM ALARM
Temperature Status Green Red
Battery Status Indicator No flashing Flashing
If the BE bit is 1 (Start/stop function active), in addition to the behavior
described above, the logging activity start can be performed in the following
way:
• Hold down the button
• Wait until the green LED lights and the red LED start blinking
• Release and press again the button within 5
seconds to confirm the
command otherwise it will be ignored.
• As command is accepted green LED will s
tay lit for 15 seconds.
• The acquisition is started.
And, if the SD
bit is 0, the logging activity stop can be performed in the fol-
lowing way:
• Hold down the button
• Wait until the green LED lights and the red LED start blinking
• Release and press again the button within 5 seconds to confirm the
command otherwise the command is ignored.
• As command is accepted red LED will stay lit for 15 seconds
.
• The acquisition is stopped.