GS3 Integrator Guide R34
For support contact us via email at support@decagon.com
or call us at 1.509.332.5600 between 7AM to 5PM PST
9
ADDRESS QUERY COMMAND (?!)
While disconnected from a bus, the address query command can be used to determine which sensor you
are currently communicating with. Sending this command over a bus will cause a bus contention where all
the sensors will respond simultaneously and corrupt the data line. This command is helpful if you are trying
to isolate a failed sensor. Here is an example output from one of our sensors (the logger command is in
bold):
?!0
PARAMETER
FIXED LENGTH
DESCRIPTION
?!
2 Chars
Request from the data logger asking for a response from any sensor
listening on the data line.
0
1 Char
Sensor Address. Returns the sensor address to the currently
connected sensor.