03
Working Principle
Air enters the compressor after filtered by the air inlet filter and is
compressed into molecular sieve tower for separation of oxygen gas and
nitrogen gas. Then, oxygen gas passes through the molecular sieve tower
and enters the fine sieve tower smoothly, while nitrogen gas is discharged
into the atmosphere through the separating valve after adsorbed by the
molecular sieve. After the concentration of oxygen gas is further increased
in the fine sieve tower, the flow rate is controlled by the flow control valve,
and then oxygen gas is wet through the humidification water tank and
finally flows through the oxygen delivery pipe for supplementary oxygen
inhalation by users.
Schematic Diagram of Oxygen Machine
Air
Filter
Separating
Valve
Oil free air compressor
Humidification
Water Tank
Cooling System
Molecular Sieve
Tower
Oxygen Gas
Control
Valve