DeFelsko PosiTest DFT User Manual

DeFelsko PosiTest DFT
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Combo: measures on all metals.
The letter “F” (ferrous, e.g. steel)
or “N” (non-ferrous, e.g.
aluminum) is displayed with
each reading depending on the
substrate. Gage automatically
recognizes the substrate.
Ferrous: measures on steel only.
The letter “F” is always displayed.
Operation: Press either button to power-up.
Place flat on surface. Hold steady. When
measurement is obtained, new reading flashes.
Lift gage or leave on the surface for repeated
measurements.
Menu: Press both buttons simultaneously
to access the gage menu. Navigate the menu
using the buttons. Press both buttons to
select:
Exit menu and return to main screen
Switch Measurement Units
Zero Calibration Adjustment: Measure the
uncoated part. Zeroing provides the best
coating thickness accuracy.
1 Point Calibration Adjustment: Also
known as ‘Offset’ or ‘Correction Value.’
Measure a known thickness. Adjust the reading
up or down using the gage buttons until
expected thickness is displayed. Press
to accept.
Coating Thickness Gage
Instruction Manual v. 4.0
mils
µm
0
1
-
+
-
+
-
+
-
+
Reset: Restore factory calibration, units,
and screen rotation settings. Handy when
an uncoated part is not available for zeroing.
Indicates that the gage is in factory
calibration
Brightness: Use or buttons to adjust
screen brightness.
Auto Rotate Lock
Power-down the gage
Average: Select to display a running
average of up to 99 readings. The
following icons will be displayed in gage menu:
Delete Last Reading: select to delete the last
reading from average. Shortcut: Press
Clear: select to clear running average.
Shortcut: Press
Off: Turn off Average
Shims: Included plastic shims provide a quick
operational check. Also used for protection on
tacky, rough or hot surfaces.
Battery Indicator: Replace AAA Alkaline
batteries when indicator turns RED.
Range: 0-40 mils 0-1000 µm
Accuracy: +/- (0.1 mils + 3%) +/- (2µm + 3%)
DeFelsko’s sole warranty, remedy and liability
are the express limited warranty, remedy, and
limited liability that are set forth on its website:
www.defelsko.com/terms
X
X
X
F
X
-
+
-
+
Question and Answer IconNeed help?

Do you have a question about the DeFelsko PosiTest DFT and is the answer not in the manual?

DeFelsko PosiTest DFT Specifications

General IconGeneral
BrandDeFelsko
ModelPosiTest DFT
CategoryMeasuring Instruments
LanguageEnglish

Summary

Gage Operation and Settings

Basic Operation

Instructions on how to power up the gage and perform measurements on a surface.

Menu Navigation

How to access the gage menu using buttons and navigate through options.

Measurement Units

Procedure to switch measurement units between mils and micrometers.

Calibration Adjustments

Details on Zero and 1 Point calibration for thickness accuracy.

Advanced Gage Functions

Reset Function

Restores factory calibration, units, and screen rotation settings.

Brightness and Rotation

Adjusting screen brightness and locking auto-rotation.

Power-Down and Averaging

How to power down the gage and manage average readings.

Average Reading Management

Options to delete last reading or clear the running average.

Accessory and Specification Information

Shims and Battery

Information on using shims for checks and battery replacement indicator.

Range and Accuracy

Specifications for the gage's measurement range and accuracy.

Related product manuals