18
DAC-515E-EN-BF
9 Technical Specifications
9.1 General
Category Specification
Size
8.5 x 11.8 x 2.7 inches (22 x 30 x 7 cm)
Weight
Approximately 3 lbs (1.4 kg) with DBP-RCx rechargeable training
battery pack
Power
Training Battery Pack (rechargeable)
Design standards
Meets applicable requirements of
• IEC 60601-1
• UL 60601-1
• 2014/35/EU Low Voltage
• IEC 60601-1-2
9.2 Environmental
Category Specification
Operating /
Maintenance
Temperature
0 – 50°C (32 – 122°F) (while operating);
0 – 40°C (32 – 104°F) (while charging
Training Battery Pack)
One Hour
Operating
Temperature Limit
(extreme cold)*
–20°C (–4°F)
Humidity 5% – 95% (non-condensing)
Standby /
Storage /
Transport
Temperature 0 – 50°C (32 – 122°F)
Humidity
5% – 95% (non-condensing)
ESD and EMI (radiated and immunity)
Refer to Chapter 10 for details
* From room temperature to temperature extreme, one hour duration.