HD2102 - 19 - V2.3
INSTRUMENT SIGNALS AND FAULTS
The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations:
Display indications Explanation
ERR
This appears if the probe has already been detected by the instrument,
but is disconnected. At the same time an intermittent beep is issued.
PROB COMM LOST
This appears if the probe has already been detected by the instrument,
but is disconnected. At the same time an intermittent beep is issued.
OVER
or
- - - -
Measurement overflow: indicates that the probe is measuring a value
exceeding the measuring range.
LOG
MEM
FULL
Memory full; the instrument cannot store further data, the memory
space is exhausted.
PROB
ERR
A probe with SICRAM module has been inserted when not admissible
for that specific instrument.
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact
the instrument's supplier.
SYS
ERR
#
Instrument management program error. Contact the instrument's
supplier and communicate the numeric code # reported by the display.
FUNC CLRD
Max, min and average values cleared.
PLS_EXIT >>> FUNC
RES_FOR_FACT
ONLY
Please exit using ESC key >>> This function is reserved to factory
calibration.
PRBE_SER #### ####
Serial number #### #### of connected probe.
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on. The
instrument issues a long beep and turns off. Replace the batteries.