22
10.2 ON LINE INSTRUMENT PROBES AND MODULES TECHNICAL DATA
10.2.1 TEMPERATURE PROBES Pt100 SENSOR USING SICRAM MODULE
Model Type Application range Accuracy
TP472I
Immersion -196°C…+500°C
±0.25°C (-196°C…+350°C)
±0.4°C (+350°C…+500°C)
TP472I.0
Immersion -50°C…+300°C ±0.25°C (-50°C…+300°C)
TP473P.I
Penetration -50°C…+400°C
±0.25°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP473P.0
Penetration -50°C…+300°C ±0.25°C
TP474C.I
Contact -50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP474C.0
Contact -50°C…+300°C ±0.3°C
TP475A.0
Air -50°C…+250°C ±0.3°C
TP472I.5
Immersion -50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP472I.10
Immersion -50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP49A.O
Immersion -70°C…+250°C ±0.25°C
TP49AC.O
Contact -70°C…+250°C ±0.25°C
TP49AP.O
Penetration -70°C…+250°C ±0.25°C
TP875.I
Globe-
thermometer
Ø 150mm
-30°C…+120°C ±0.25°C
TP876.I
Globe-
thermometer
Ø 50mm
-30°C…+120°C ±0.25°C
TP87.O
Immersion -50°C…+200°C ±0.25°C
TP878.O
TP878.1.O
Solar panels +4°C…+85°C ±0.25°C
TP879.O
For compost -20°C…+120°C ±0.25°C
Common characteristics
Temperature drift @20°C 0.003%/°C
Note: probes identified by the last letter “I” in the ordering code are made with wire wound sensor,
probes identified by the last letter “O” are made with thin film sensor.