C
Cynthia PerezAug 30, 2025
Why won't the boost power mode activate on my Dentsply Sirona Cavitron 300?
- SSherri SmithAug 30, 2025
If the Dentsply Sirona Dental equipment Boost Power mode does not activate and the Diagnostic Display does not indicate “Boost” mode, it could be due to: * The Foot Control not being fully depressed. Make sure to depress the Foot Control fully. * A defective Foot Control. Replace the Foot Control.