18
Overview
Specifications
Model MXX-123 MXX-612 MXX-412 MXX-212
Weighing capacity g 120 610 410 210
Readability g 0.001 0.01 0.01 0.01
Tare range (subtractive) g 120 610 410 210
Linearity ±g 0.003 0.03 0.03 0.03
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Response time (average) s 2.5 2 2 2
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update:
0.1–0.8 (depends on filter level selected)
Calibration weight g 100 (F1) 200 (F2) 200 (F2) 200 (M1)
Net weight, approx. kg 1.2 1.35 1.35 1.2
Pan size mm 97 d 142+130 142+130 97 d
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 0.75 0.75 0.75
Hours of operation w/ 9V battery:
Alkaline (approx.) h 11 11 14
Rechargeable, fully ch.
(NiMH), avg. h 2.5 2.5 4
Model MXX-5001 MXX-2001 MXX-601 MXX-10 MXX-5
Weighing capacity g 5000 2000 600 10000 5000
Readability g 0.1 0.1 0.1 1 1
Tare range (subtractive) g 5000 2000 600 10000 5000
Linearity ±g 0.2 0.2 0.2 2 2
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Response time (average) s 2 2 1.5 1.5 1.5
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update:
0.1–0.8 (depends on filter level selected)
Calibration weight kg 5 (F2) 2 (F2) 0.2 (M2) 5 (M1) 5 (M2)
Net weight, approx. kg 1.1 1.1 1.25 1.1 1.1
Pan size mm 142+130
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 1 0.75 0.75 0.75
Hours of operation w/ 9V battery:
Alkaline (approx.) h 11 11 14 14 14
Rechargeable, fully ch.
(NiMH), avg. h 2.5 2.5 4 4 4

Table of Contents