TROUBLESHOOTING
Issue Possible Cause Remedy
Water condensation collecting in the
hose causing a gurgling sound when
using optional humidier.
1. Humidication is too high. 1. Reduce humidier heater setting.
2. Room temperature uctuates from higher to lower levels
through the night.
2. Increase room temperature.
Heater setting is not visible on display. 1. The ow generator is not fully seated onto the humidier cradle.
1. Refer to System Assembly to ensure proper contact is
made.
SPECIFICATIONS
Humidifier
Size with CPAP Device: ..................................................................................................................................................... 6.3” (16.1 cm) H x 6.1” (15.5 cm) W x 7.6” (19.3 cm) D
Weight: .................................................................................................................................................................................................................. 3.8 lbs (1.75 kg) w/CPAP device
Humidity Output (in the operating flow range): ................................................................................................................................................................................≥10 mgH
2
O/l air
Maximum Power Consumption: ................................................................................................................................................................................................ 40 watts (humidifier)
Typical Power Consumption with CPAP ....................................................................................................................................................................................................... 25 watts
DV6HH Electrical Requirements: ........................................................................................................................................................................... 11-17 VDC powered from CPAP
Heater Plate Thermal Cutout:.............................................................................................................................................................................................................. 239˚F (115˚C)
Power/Temperature Control: ........................................................................................................................................................ Humidity Setting of 1 (minimum) to 5 (maximum)
Heater Plate Temperature: ............................................................................................................................................................................................91˚F to 149˚F (33˚C to 65˚C
Operating Temperature Range: ......................................................................................................................................................................................41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range: ................................................................................................................................................................................. 15% to 93% R.H. non-condensing
Operating Atmospheric Pressure Range: ....................................................................................................................................................................................... 70.0 – 106.0 kPA
Operating Flow Range: .................................................................................................................................................................................................................. 15 L/M to 60 L/M
Not manufactured with natural rubber latex.
MAINTENANCE
Humidier Water Chamber (single-patient use) - Clean the
chamber daily. The chamber should be replaced according to the
schedule recommended by your provider. Inspect the chamber
monthly; replace if damaged.
Humidier Cradle - No maintenance needed for humidier cradle.
EXPECTED SERVICE LIFE
Humidier Cradle - 5 years
Humidier PulseDose Valve (only model DV6HHPD) - 5 years
Humidier Water Chamber - 1 year
PRODUCT DISPOSAL
The outer packaging is made of environmentally friendly materials that can be used as secondary
raw materials. If you no longer need this packaging, bring it to your local recycling and waste
disposal facility according to the applicable regulations.
The equipment, including accessories and internal components, does not belong in your regular
household waste; such equipment is manufactured from high-grade materials and can be recycled
and reused. The PC board, heater plate, and wiring should be removed from the unit and recycled
as electronic waste. The remaining plastic components should be recycled as plastics.
The European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE)
requires that electrical and electronic equipment be collected and disposed of separately from
other unsorted municipal waste, with the aim of recycling it. The crossed out waste bin symbol
indicates that separate collection is required.