7.2 Starting up the device
!
I
At the first starting up of the device or after switching it off for a long time the
display of the digital thermometer (fig. 2, page 4) shows the current ambient
temperature alternating with the error message “E6”.
The alarm LED will light up but no audible alarm will sound.
As soon as the inner temperature reaches the operating temperature range
(2 °C 8 °C) the current temperature is shown.
Reset the thermometer to avoid a wrong maximum temperature value.
Press the button for 5 seconds.
7.3 Display backlight in battery mode
The thermometer display has a blue backlight to make the temperature values easier
to read. The display backlight works continuously while the device is connected to
the mains.
In battery mode the display backlight will switch off after 2 minutes.
To activate the display backlight for 3 seconds, briefly press the button.
7.4 Temperature values
The thermometer stores the values of the minimum and maximum inner temperature
since the last reset.
To reset these values:
Press the button for 5 seconds.
CAUTION!
Do not store medicines and medicinal products before the E6 error
code has disappeared and the operating temperature range has been
reached.
NOTE
The temperature values are shown in °C. This cannot be changed.