Using the refrigerator HC302D, HC502D
22
7. 5 A la r m s
A
There are five special cases when alarms occur:
The alarm can be muted and unmuted:
Press the button.
I
NOTICE!
If you are not able to eliminate the problem triggering the alarm,
store the medicinal products somewhere else as soon as possible.
In case of power failure, the inner temperature will raise in 2 hours
(hold over time) from 2 °C to 10 °C at an ambient temperature
of 25 °C.
Alarm event Number of beeps Duration of the beeps
Temperature out of range 1 beep / 5 sec 1 sec
Power failure 1 beep / 3 sec 0.5 sec
Inner fan is blocked n.a. permanent beeping
Door open for > 20 sec
1 beep / 2 sec 1 sec
Connection problem between
display and controller
Door open for > 60 sec 2 beeps / 2 sec 0,5 sec
NOTE
If another alarm is triggered the alarm signal is automatically
unmuted.
After all alarm events have passed, a muted alarm signal must be
manually reset to unmuted.