Verification of the signal coverage and strength
The signal strength shall always be verified during, or in association with,
the installation/deployment of the CareMobile in order to ensure trouble-
free operation.
The verification is done as follows: Press the yellow and green button si-
multaneously, then release. The three indicators showing the current sig-
nal strength during four seconds as follows:
+
Lights red
The measured signal strength 0
Not defined
+
Lights green
The measured signal strength 14 or lower
Not sufficient
!
+
Lights red
Lights green
The measured signal strength 15 or higher
Good
!
!
+
Lights green
Lights red
Lights green
The measured signal strength 20 or higher
Very good
The higher the value the higher the strength. You need to get at least
strength Good for good communication. If this can’t be achieved, try to
move the unit in different locations in the residence until approved value
are fulfilled.
Note! If accepted value can’t be achieved an external antenna must be
used.