42
Remote Operation
7.4.3 *ESE?
Syntax
*ESE?
Description
This query allows the user to determine the current contents of Standard Event
Status enable register.
The value returned corresponds to the binary-weighted sum of all bits enabled
by the *ESE command.
Note: This command can NOT be concatenated with other commands. It must
be issued as a single command.
7.4.4 *ESR?
Syntax
*ESR?
Description
This query allows the user to determine the current contents of Event Status
register. Reading the Event Status Register clears it.
The status is returned as a decimal value which corresponds to the binary-
weighted sum of all bits set in the register.
Note: This command can NOT be concatenated with other commands. It must
be issued as a single command.