Do you have a question about the Dräger Alcotest 7410Plus RS and is the answer not in the manual?
Repair of the instrument may only be carried out by trained personnel. Use original Dräger spare parts.
Use a new mouthpiece, no alcohol/solvent vapors, keep distance from phones/transmitters, keep vent open.
15 min interval after alcohol intake, no residues from drinks/sprays, breathe evenly, avoid deep breaths.
Follow simulator instructions. Insert adapter, attach clip, switch on. Display shows 'CAL'. Green Ready-light and tone indicate readiness.
Specifies calibration interval depends on required accuracy, typically 6 months.
Defines the permissible calibration range for measuring units like %0, mg/L, g/L.
Provides standard calibration values using Dräger Ethanol-Standard solution.
Covers instrument switching off, yellow light issues, no ready light, no tone, and specific error messages like E 0, E 1.
| Brand | Dräger |
|---|---|
| Model | Alcotest 7410Plus RS |
| Category | Test Equipment |
| Language | English |