EasyManuals Logo
Home>Dräger>Test Equipment>Alcotest 7410Plus RS

Dräger Alcotest 7410Plus RS User Manual

Dräger Alcotest 7410Plus RS
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
32
Fault message in display Cause Remedy
E 2 Instrument does not
recognize calibration clip
Attach calibration clip correctly
E 3 Calibration not sucessful
Worn sensor
Calibrate again, by trained service
personnel, e.g. by DrägerService
Sensor must be replaced by DrägerService
E 4 Calibration value not permissible Calibrate with permissible value
E 5 Fault in sampling system Repeat measurement,
if fault message re-appears, call DrägerSer-
vice
E 6 Instrument malfunction Call DrägerService
E 7 Data memory corrupted Call DrägerService
FCC followed by Err
red light + sound signal
Harmful electromagnetic interference was
detected
Redress interference,
repeat measurement.
000 followed by dAt
red light + sound signal (only Alcotest
7410
Plus
RS)
Memory completely occupied,
data area corrupted or not initialized
Readout of stored data. Clear memory after-
wards using PC software "VIEW 2000".
-LI red light + sound signal Low voltage of data storage battery Readout of stored data using PC software
"VIEW 2000" (only Alcotest
7410
Plus
RS).
Call DrägerService.
dS red light + sound signal (only if display of
time is activated)
Calibration interval expired Calibrate the instrument, by trained service
personnel, e.g. DrägerService.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Alcotest 7410Plus RS and is the answer not in the manual?

Dräger Alcotest 7410Plus RS Specifications

General IconGeneral
BrandDräger
ModelAlcotest 7410Plus RS
CategoryTest Equipment
LanguageEnglish

Related product manuals