EasyManua.ls Logo

Dräger Alcotest 6510 User Manual

Dräger Alcotest 6510
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
D
Alkoholimittari ammatti- ja viranomaiskäyttöön
Breath Alcohol Monitor
Alcotest
®
6510
ST-207-2004_sw.eps
Käyttöohje nro 9023757 -- A.023
Instructions for Use

Table of Contents

Questions and Answers:

Dräger Alcotest 6510 Specifications

General IconGeneral
TypeBreath Alcohol Tester
Resolution0.01 mg/L
Sensor TypeElectrochemical Sensor
Operating Temperature-5 °C to +50 °C
Storage Temperature-20 °C to +60 °C
Humidity Range10% to 95% RH (non-condensing)
DisplayLCD
Power SupplyRechargeable battery or AC adapter
InterfaceUSB
CalibrationAutomatic
Calibration Interval6 months

Summary

Safety Guidelines for Drager Alcotest 6510

Strictly Follow Instructions

Emphasizes understanding and strict observation of instructions for safe operation.

Expert Maintenance Required

Maintenance must only be performed by experts using authentic spare parts.

Liability for Proper Function and Damage

Liability transfers if serviced by unauthorized personnel or used improperly.

Intended Use and Preparation

Intended Use of Alcotest 6510

Device for measuring alcohol concentration in breath for professional/official use.

Preparation and Initial Setup

Caution: Charging Batteries in Instrument

Warning: Charging batteries while in the instrument will cause damage.

Important Notes on Symbols

Warnings about service socket and not inserting power cables.

Performing a Measurement

Requirements for the Person Tested

Waiting periods, avoiding mouth alcohol, proper breathing, and minimum air volume.

Automatic Measurement Process

Steps before automatic measurement: insert mouthpiece, switch on device.

Measurement Sequence Displays

Describes the 'WAIT' and 'READY' displays during the measurement process.

Breathing and Analyzing Sample

Instructions for the test subject to blow evenly and continuously.

Waiting for Analysis

Indicates the device is analyzing the sample.

Displaying the Test Result

Explains when and how the measurement result is displayed, including units.

Handling Breath Sample Errors

Insufficient Breath Volume Error

Describes error for low breath volume and how to repeat.

Breath Interruption Error

Describes error for improper exhalation and how to repeat.

Manual Sampling Procedure

Device Settings Configuration

Measurement Parameters

Measurement Sample Modes

Measurement Criteria and Settings

Calibration Procedures

Performing Calibration

Calibration Process Steps

Steps for performing calibration using setpoint, gas type, and supply gas.

Accuracy Check After Calibration

Accuracy Check Procedure

How to perform an accuracy check after calibration using calibration gas.

Error Message Handling

Internal System Error Display

Displays internal error messages and advises contacting DrägerService if the error persists.

Maintenance and Storage

Cleaning and Battery Replacement

Battery Disposal Warnings

Warnings about battery disposal, fire, recharging, and explosion risks.

Calibration with Simulators and Gas

Calibration Using Mark II Simulator

Steps for calibrating with the Mark II simulator, including connection advice.

Calibration Using Dry Gas

Instructions for calibrating using dry gas and a hose.

Calibration Procedure and Accuracy Check

Calibration and Accuracy Check Steps

Describes the calibration procedure and subsequent accuracy check.

Troubleshooting Faults and Errors

Instrument Auto Shut-off Fault

Possible cause (battery) and remedy (replace batteries).

Instrument Not Switching On Fault

Possible cause (battery) and remedy (replace batteries).

Insufficient Breath Volume Fault

Possible cause (weak blow) and remedy (ask to blow harder).

Calibration Error Message

Error message and remedy (recalibrate by expert).

Accuracy Check Error Message

Error message and remedy (repeat accuracy check, call service).

Internal System Error Message

Error message and remedy (remove/reinsert batteries, contact service).

Troubleshooting: Faults, Causes, and Remedies

Instrument Switches Off Automatically

Cause: batteries discharged. Remedy: replace batteries.

Instrument Cannot Be Switched On

Cause: batteries discharged. Remedy: replace batteries.

Insufficient Volume or Blowing Interruption

Cause: testperson not blowing hard/evenly. Remedy: ask to blow harder/more evenly.

Calibration Error

Cause: Calibration not successful. Remedy: Calibrate again by trained service personnel.

Accuracy Check Error

Cause: Accuracy check not successful. Remedy: Repeat accuracy check, call DrägerService.

Internal System Error

Cause: Instrument error. Remedy: Remove/insert batteries, call DrägerService.

Technical Data Specifications

Related product manuals