Safety Features
10.9 System Failure
If the system no longer responds to an action:
1. Ventilate the patient manually
.
2. Switch the machine off and on again
.
3. Cancel the self test.
4. Check the vaporizer setting.
5. Close all flow controls (except O2), and run 100% O2.
If the breathing bag does not fill
with fresh gas or manual ventilation is not
possible:
• check the oxygen supply; open cylinder valves if necessary
If fresh gas is still not delivered
:
• close all flow controls
• disconnect patient
from workstation and use another device
• turn workstation off