8 Supplement Paper Apollo SW 3.2n, SW 4.n
Chapter 9 - Alarms
Adjustment Range of the Alarm Limits
during Operation
With Software 3.2n:
With Software 4.n:
Chapter 11 – Cleaning and Maintenance
Maintenance intervals
Refer to the Apollo Instructions for Use.
Chapter 13 – Specifications
CO2 Module
FiCO2
[Vol.%]
0.1 to 1.4 Vol.%, – –
FiCO2
[Vol.%]
0.1 to 3.5
1)
, – –
1) DrägerService can change the upper limit to 1.4.
WARNING
Risk of patient injury.
To prevent the CO
2 module from malfunction-
ing, only the Dräger pressure reducer CO2 Pin
index (part no. 8607429) should be used. Only
use a CO
2 cylinder with this pressure reducer
to supply the CO2 module.
Inlet connection: Pin-Index coded pressure regulator
Inlet pressure 5 kPa x 100 ±20 %
(72.5psi ±20%)
CO
2 flow dosing: 0 to 0.6 L/min ±10 %
with an inlet pressure of 5 kPa x 100
(72.5 psi)
CO
2 flow meter 100 to 1000 mL/min
Accuracy ±50 mL/min
Resolution 50 mL/min