D-12377-2016
Disassembly/Assembly Muffler
Open the flap by lifting the lower
edge upwards.
Close the flap.
Pull the muffler as far it will go
to the left.
Push the muffler to the
right into the socket.
Remove the expiratory valve from
the fitting.
Turn the locking ring as far as
possible to the left.
Remove and empty the water
trap container.
Fit the muffler into the socket.
Remove the muffler from
the socket.
Remove the diaphragm.
91 03 108 | 17.03-2 | AW | LE | Subject to modifications | © 2016 Drägerwerk AG & Co. KGaA
Fit the water trap container.
Fit the diaphragm onto the edge
of the expiratory valve housing.
Turn the locking ring as
far as it will go to the right
until it clicks into place.
Turn the locking ring as far as
possible to the left and push the
expiratory valve into the fitting.
Dräger Babylog
®
VN500
Neonatal Expiratory Valve
and Neonatal Flow Sensor
Disassembly/Assembly and Reprocessing
Disassembly/Assembly Neonatal Expiratory Valve
The muffler is only attached to the Babylog
®
VN500. It is only available with
an 11 mm diameter port. The Babylog
®
VN500 is dedicated to neonatal and
pediatric patient population.
This guideline does not replace or amend the instructions for use for Infinity Acute Care System Workstation Neonatal Care.
The instructions for use and the hospital guidelines and policies always have to be observed.

Other manuals for Dräger Babylog VN500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Babylog VN500 and is the answer not in the manual?

Dräger Babylog VN500 Specifications

General IconGeneral
BrandDräger
ModelBabylog VN500
CategoryMedical Equipment
LanguageEnglish

Related product manuals