Troubleshooting
530 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
!
(Any
SpO2
MCable)
SpO2 H/W failure Parameter values
are replaced by ***
Hardware failure Check for faulty
MCable
Contact
Dräger-
authorized
service
personnel.
! SpO2 Interference
Detected
2)
Interference
detected
Parameter values
are replaced by ***
Interference such as
artifact or too much
ambient light was
detected.
–Make sure the
sensor is
properly
attached.
–Make sure that
no nail polish or
some other
substance is
blocking the
light.
Change the
sensor location.
1)
%0 is a placeholder for the parameter label SpO2.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
3)
In the parameter field the parameter value is replaced by ***
Priority Message Parameter field Problem Solution