NIBP Problems
NIBP did not occur as
desired
Monitor has message
of NIBP open line
Monitor shows mean
b/p only
NBP Low inflation
Limit
NBP Cannot Measure
Possible Cause
Interval change not
saved.
Leak between hose
and cuff.
Incorrect cuff size
Patient’s pulse is too
low for monitor to
derive systolic and
diastolic pressure
values but large
enough to report mean
pressure.
Patient’s systolic
pressure is higher than
the maximum allowed
inflation limit.
Pulse profile is too
poor to make a reliable
measurement (usually
owing to persistent
motion artifact).
Action
Initiate b/p or double
click start/stop nbp
with new interval
Disconnect and
reconnect hose to
cuff.
Ensure that cuff size
and placement is
correct. Check the
patient and treat if
necessary. Check
hose and cuff.
Select the next higher
NBP inflation limit
setting.
Check the patient and
treat if necessary.
Move cuff to a limb
with less movement.
Restart measurement.
If message persists
contact Biomed or
Technical Support.

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals