EasyManuals Logo

Dräger MMG 0001 User Manual

Dräger MMG 0001
Go to English
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56 Technical Manual | X-act
®
7000
en | Troubleshooting
Optics - error 2097193 The camera could not find
any MicroTubes
Check whether Micro-
Tubes were completely
drawn into the device; if
necessary, push the
MicroTubes in a bit faster
and with slightly more
pressure; check whether
all orange pins of the
MicroTubes have already
been depressed; if neces-
sary, use new MicroTubes
Pneumatics - blockage
detected
Blockage of the pneumatic
system detected
Make sure that the gas
inlet or outlet is not sealed
(e.g. by fingers, contami-
nation); if the blockage is
not promptly removed, the
X-act 7000 reports a
“Pneumatics - error”
RFID - multiple RFIDs Several MicroTubes are
located in the immediate
vicinity of the X-act 7000
and the associated RFID
tags are simultaneously
detected
Remove all MicroTubes,
except for the one to be
measured, from the imme-
diate vicinity of the X-act
7000 and repeat the pro-
cess
RFID - writing failed The number of remaining
measurements on Micro-
Tubes could not be written
back onto the Microtubes’
RFID tag
As a result, X-act 7000
displays one remaining
measurement too many;
the correct number of
remaining measurements
can be read out based on
the orange pins that have
not been depressed on the
MicroTubes
RFID - data corrupted Information on the RFID
tag of the inserted Micro-
Tubes is faulty
MicroTubes can no longer
be used; insert new Micro-
Tubes for further measure-
ments
Remote pump - not found Communication with the
X-am pump could not be
established
Check whether the X-am
pump is ready for opera-
tion; charge if necessary;
make sure that the coupler
and X-am pump are cor-
rectly fitted on the X-act
7000; make sure that the
infrared interfaces on the
X-act 7000 and X-am
pump have not been con-
taminated; make sure that
no foreign bodies are
located in the coupler
Warnings Cause Remedy