EasyManuals Logo

Dräger Oxylog 2000 plus Supplement To The Instructions For Use

Dräger Oxylog 2000 plus
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
Supplement to the instructions for use Oxylog 3000 plus 3
English
Problem solving
Messages in the alarm message field
Technical data
Device specifications
Alarm Cause Remedy Rank
! CO
2 inac-
curate or
failed
The CO
2 measure-
ment is inaccurate.
Use external CO2
monitoring.
Contact
DrägerService.
42
The CO
2 sensor is
defective.
Replace the faulty
CO
2 sensor.
The zero calibra-
tion or filter check
of the CO
2 sensor
was activated and
no valid measured
values were gener-
ated.
Wait until the CO
2
zero calibration/fil-
ter check is com-
pleted.
!! etCO
2
low
The lower alarm
limit for the end-
tidal CO
2 concen-
tration has been
exceeded.
Check patient con-
dition.
27
Check alarm limits.
Adjust the alarm
limit, if necessary.
The CO
2 sensor
has fallen off the
cuvette.
Reattach the CO2
sensor to the
cuvette.
! No int.
battery
charging
Internal battery
cannot be charged
due to a faulty bat-
tery or too hot or
cold environment.
Press the Alarm
Reset key to con-
firm the alarm.
Remove the inter-
nal battery and
reinsert or replace
it.
44
Protection class
against ingress of
liquids
IP34, the device is splash-proof from all
directions and protected against foreign
objects >2.5 mm (0.1 in)

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals