Supplement Oxylog family 3
English
Supplement Oxylog family en English
Supplement to the instructions for use
This supplement applies to the following devices:
– Oxylog 2000 plus
– Oxylog 3000
– Oxylog 3000 plus
– Oxylog VE300
Keep this supplement with the instructions for use.
This supplement updates the information of the instructions for
use in the following chapters.
Maintenance
Maintenance schedule
WARNING
To properly use this medical device, read and comply with
the instructions for use and this supplement.
Component Interval Task Personnel responsible
Dust filter Every 2 years Replace
1)
Service personnel
Internal rechargeable battery – Every 4 years
– When the battery no longer
remains charged for the
specified operating time
2)
Replace Users
Workstation Every 2 years Inspection and preventive main-
tenance
Service personnel
1) The dust filter can be treated as household waste.
2) Refer to "Technical Data" section in the corresponding instructions for use for battery operating time.