Testing
64
Проверка
Checking the Relief Valve
1 Open shut-off valve at the test unit.
2 Pump slowly until
3 the relief valve opens (flow sound).
4 The opening pressure indicated at
the test unit should amount to 4 to
7 mbar.
Disconnect pump from the test unit.
The breathing circuit is de-aereated
automatically.
The low-pressure warning generated by
Bodyguard during this test is irrelevant
here.
High-Pressure Leak Test
1 Set "Leak test".
2 Open the cylinder valve.
3 Watch the Bodyguard: the cylinder
pressure is indicated here.
If cylinder pressure is lower than
165 bar:
Charge the oxygen cylinder.
15821726.eps
4
2
1
3
15921726.eps
1
3
2
Проверка избыточного
клапана
1 Откройте отсечной клапан на
испытательном устройстве.
2 Медленно создавайте избыточное
давление, пока
3 не откроется избыточный клапан
(слышен звук вытекающего
воздуха).
4 Открывающее давление,
показанное на испытательном
устройстве, должно лежать между
4 и 7 мбар.
Отсоедините насос от
испытательного устройства.
Давление в дыхательном контуре
будет автоматически сброшено.
Предупреждение о низком давлении,
которое выдается Bodyguard в ходе
этой проверки, не имеет значения.
Проверка утечек из
высоконапорных линий
1 Установите режим "Проверка
герметичности".
2 Откройте вентиль баллона.
3 Наблюдайте показания Bodyguard:
он показывает давление в
баллоне.
Если давление в баллоне ниже
165 бар:
Заправьте кислородный баллон.