Savina Alarm management
Display of alarms
13
Alarms are signaled optically and acoustically according to their alarm
priority.
1. Optical alarm signals is displayed in the alarm message field.
For alarms with high priority, the red LED flashes.
For alarms with medium priority, the yellow LED flashes.
For alarms with low priority, the yellow LED lights up.
2. Acoustic alarm signals
The alarm with the highest priority is signaled acoustically. The alarm
signal continues to sound until either the cause for the alarm has
been resolved or the alarm signal is suppressed.
Alarm display Warning Caution Note
Priority Top !!! Medium ! ! Low !
Audio length
5 tone seq.
sound twice
3 tone seq.
sound once
2 tone seq.
sound once
Audio alert Every 7 sec. Every 20 sec. ---
All user set alarms on single page
V
Ti
alarm only in mask ventilation
Alarm limits for FiO
2
when using high pressure inlet is automatically set
1
1

Other manuals for Dräger Savina

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Savina and is the answer not in the manual?

Dräger Savina Specifications

General IconGeneral
FiO221 - 100 %
Ventilation ModesCPAP/ASB
DisplayColor
Power Supply100 – 240 V AC, 50/60 Hz
Patient RangeAdult, Pediatric, Neonatal
Tidal Volume Range20 to 2000 mL
Screen Size12.1 inch
Inspiratory Time0.1 - 10 s

Related product manuals