Page #27 background image
© Drägerwerk AG & Co. KGaA, 2019
Savina Cleaning, disinfection, and
sterilization
27
Safety information on reprocessing
WARNING
Risk of infection or damage to device
Clean, disinfect, and sterilize the device and accessories using validated processes.
CAUTION
Risk of infection
If reusable medical devices are not reprocessed, there is an increased risk of infection to both hospital staff and patients.
Clean and disinfect reusable medical devices after each use.
Protective clothing, eye protection etc. must be worn.
Observe the hospital hygiene regulations!
Reprocess the device after every patient.
Recommendation:
Change the breathing circuit and expiration valve every 24 hours. Keep replacement systems ready.
CAUTION
The screen is made of Plexiglas. Do not treat with alcohol or agents containing alcohol. Danger of cracking.

Other manuals for Dräger Savina

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Savina and is the answer not in the manual?

Dräger Savina Specifications

General IconGeneral
FiO221 - 100 %
Ventilation ModesCPAP/ASB
DisplayColor
Power Supply100 – 240 V AC, 50/60 Hz
Patient RangeAdult, Pediatric, Neonatal
Tidal Volume Range20 to 2000 mL
Screen Size12.1 inch
Inspiratory Time0.1 - 10 s