Savina Reprocessing list
32
Applicable to non-infectious patients.
The reprocessing list contains approximate values only. The instructions of the hospital’s infection control officer responsible have priority.
1) Keep spring-loaded valves (water trap, pneumatic medication nebulizer) open during reprocessing.
2) Nebulization may lead to increased deposits making it necessary to exchange the parts more often.
3) For additional information, see corresponding instructions for use: “Reprocessing the breathing gas temperature sensor manually”.
Items which can be reprocessed
Recommended
reprocessing intervals
Pre-cleaning
Machine cleaning and
disinfection
Manual
Sterilizatiom
Cleaning Disinfection
Breathing hoses
Per patient/weekly
Yes Yes
According to the corresponding instructions
for use
Yes
Y
-piece
Per patient/weekly
Yes Yes Possible Possible Yes
Water traps
1)
Water trap container
Expiratory valve
Per patient/weekly
2)
Yes Yes Possible Possible Yes
Diaphragm
Flow sensor sleeve
Water trap container
Breathing gas temperature sensor
Per patient/weekly
No No Outside
3)
Outside
3)
Yes
Flow sensor
Per patient/weekly
2)
According to the corresponding instructions for use
Breathing gas humidifier
Per patient/weekly
According to the corresponding instructions for use
Medication nebulizer
1)
According to the corresponding instructions for use
Parts for adapting
Per patient/weekly

Other manuals for Dräger Savina

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Savina and is the answer not in the manual?

Dräger Savina Specifications

General IconGeneral
FiO221 - 100 %
Ventilation ModesCPAP/ASB
DisplayColor
Power Supply100 – 240 V AC, 50/60 Hz
Patient RangeAdult, Pediatric, Neonatal
Tidal Volume Range20 to 2000 mL
Screen Size12.1 inch
Inspiratory Time0.1 - 10 s