Trouble shooting Bellavita handset
Trouble shooting Bellavita Handset Stand 11/2009 Page 1 of 2 / HJR
Failure
Possible causes
Trouble shooting
Led
didn’t
A
cc
upack
unplugged
from
board inside
Replace handset
LED
is flashing
,
but bathlift
didnt move.
Coiled cable overstretched
and thereby unplugged
Replace handset
LED
twinkle green
during
pressing the UP-button, but
the bathlift don`t move
Hands
et is not connected
correct Alert function
Switch off handset
,
connect
correct and switch on again.
LED
twinkles
r
ed
(
amber
)
in
spite of fully accu.
Board
in
a wrong
m
od
e
Switch off handset and
disconnect, the connect again
and switch on.
Here you can also do the
selftest in case of insecurity,
the test must always start with
a green flashing LED !
The handset should always be switched off before connecting
(
as well as for
use at the
bathlift or charging). !
Whil
e
c
h
arging the LED
twinkles red
Board
in a wrong
m
od
e
Disconnect
power supply
,
Switch on handset- and then
switch off. Connect now first
handset and power supply
and then connect power
supply with the socket.
While
c
h
arging the LED
twinkles green
Board
in a wrong
m
od
e
Se
e above
LED
didn`t flash while
charging
P
ower supply
defective or bad
connected
Check connection
,
if
necessary replace power
supply.
Note:
If there are still problems with the charging mostly a reset will help. To make a
reset, please do twice the selftest but without the relais test, this means switch off
the handset when the red Led is flashing !
Requirements fort he selftest: The accu`s must be charged so that the LED shows green or
amber, not red ! Also the handset should not be connected neither with the bathlift nor with
the power supply.
How to do the selftest, see next page:
Question and Answer IconNeed help?

Do you have a question about the Drive Bellavita handset and is the answer not in the manual?

Drive Bellavita handset Specifications

General IconGeneral
BrandDrive
ModelBellavita handset
CategoryMedical Equipment
LanguageEnglish

Summary

Trouble shooting Bellavita Handset

Trouble shooting Notes

Important notes for troubleshooting, including reset procedures and selftest requirements.

Selftest Procedure

Step-by-step guide on how to perform a selftest on the Bellavita hand control.

Related product manuals