SECTION III
MAINTENANCE
3.1. BATTERY LIFE
3.1.1 The test set battery should be replaced when its load voltage has dropped
to 7 volts or the display reads LB when the test set is turned on.
3.2. BATTERY REPLACEMENT
3.2.1 Any 9-volt battery may be used, but longer service will be obtained by
the use of premium batteries such as NEDA Type 1604A. The battery is ac-
cessible in the Test Set by removing the battery access door on the bottom rear
of the case and dropping out the old battery. A snap type connector facilitates
change of battery and prevents reversal of polarity, but as an additional pre-
caution, always turn the test set off before making the battery change.
3.3. TEST SET CALIBRATION
3.3.1 The TS200 Test Set is calibrated at the factory during manufacture.
Under normal operating conditions the test set will not require calibration for
a period of one year. The calibration due date is shown on the back of the test
set.
3.3.2 For calibration procedures please contact Dukane Corporation or
refer to the calibration procedures in the CT200 manual.
03TM-0045 REV 00