MSP1
 and 
MSP2
 (Maintenance Set Point 1 & 2) SUB MENUS
Adjust for the desired maintenance set points when the unit is placed in the 
maintenance mode. The deadband is fixed at 2% of full scale. To enter or 
leave the maintenance mode, press and hold the    for 8 seconds.
ADDR
 (Modbus
®
 Protocol Address ) SUB MENU
Modbus
®
communication instrument address. Set from 1 to 247. This num-
ber must match the address number used by the host computer. 
To obtain the Digihelic
®
Controller Modbus
®
register list please visit 
www.dwyer-inst.com
See page 4 for wiring diagram.
WR
 (Modbus
®
 Write Protect) SUB MENU
d.S
- Disables write commands from Modbus
®
. Protocol can only read 
information from the instrument.
En
- Enable write commands from Modbus
® 
Protocol. Modbus
®
can read 
information from and write information to the instrument. 
ZERO
 and 
SPAN
 (Calibration of Zero and Span) SUB MENUS
The lower display reads 
CAL
in this mode.
ZERO
 Calibration
NOTE: For accurate calibration, DO NOT apply any pressure when 
performing this function.
With the display reading 
ZERO
, press the ENTER key. The upper display will 
blink. Press ENTER again to complete the zeroing of the instrument or press 
the 
MENU
key to cancel.
SPAN
 Calibration
With the display set to 
SPAN
, apply full scale pressure to the unit. Press the 
ENTER key. The upper display will blink. Press ENTER again to complete the 
calibration or press the 
MENU
key to cancel.
SP/ALSP/AL
MENU
E
RST
PAGE 1 OF 12
8/25/03
©Copyright 2009 Dwyer Instruments, Inc. 
Printed in U.S.A. 7/09
FR# 443246-00 Rev. 4
DWYER INSTRUMENTS, INC. 
Phone: 219/879-8000     www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, IN 46361, U.S.A.          Fax: 219/872-9057          e-mail: info@dwyer-inst.com
Modbus
®
is a registered trademark of Schnieder Automation and Company.
B-32:B-32  7/9/09  2:52 PM  Page 20
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30