Long-Term Care Bed • D100 (Reorder No. 12001) 3
DECLARATION – ELECTROMAGNETIC EMISSIONS
Radio Frequency Interference (RFI) inuences most electronic
equipment. Caution should be exercised regarding the use of
portable communications equipment in the area around such
equipment. If RFI causes erratic behavior, shut the bed off
immediately. Leave it off while the transmission is in progress.
Interference to electronic equipment may occur
in the vicinity of devices marked with this symbol:
Information about Electromagnetic Emissions
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The care bed is intended for use in the electromagnetic environment specied below. The customer or user of the care bed
should ensure that it is used in such an environment.
Emitted interference Compliance Electromagnetic Environment -
Guidelines
RF emissions according to CISPR11 Group 1 The care bed uses RF energy only for
its internal functioning. Therefore the
RF emissions are very low and it is
unlikely that nearby electronic devices
will be disturbed.
RF emissions according to CISPR11 Class B The care bed is designed for use in
all establishments including domestic
establishments and those determined to
be directly connected to a public supply
network that supplies buildings used for
residential purposes.
Emissions of harmonics according to
CISPR11
Class A
Emissions of voltage uctuations/Flicker
according to IEC 61000-3-3
Complies
Information about Electromagnetic Interference Immunity
Guidance and Manufacturer’s Declarations - Electromagnetic Interference Immunity
The care bed is intended for use in the electromagnetic environment specied below. The customer or user of the care bed
should ensure that it is used in such an environment.
Interference Immunity
Certication
IEC 60601 Test Level Compliance Level Electromagnetic Environ-
ment - Guidelines
Electrostatic Discharge (ESD)
according to IEC 61000-4-2
± 6 kV Contact discharge
± 8 kV Air discharge
± 6 kV Contact discharge
± 8 kV Air discharge
Floors should be wood, con-
crete or ceramic tile oors.
If the oor is covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical Fast Transients/Bursts
according to IEC 61000-4-4
± 2 kV For power lines
± 1 kV for input and output
lines
± 2 kV For power lines
± 1 kV For input and output
lines
The quality of the supply
voltage should be equivalent
to that of a typical business
or hospital environment.
Surges according to IEC 61000-
4-5
± 1 Kv Voltage phase-
phase conductor
1 Kv Voltage phase-
ground conductor
± 1 Kv Voltage phase-phase
conductor
1 Kv Voltage phase-ground
conductor
The quality of the supply
voltage should be equivalent
to that of a typical business
or hospital environment.
RF emissions according to CIS-
PR11
Class B The care bed is designed
for use in all establishments
including domestic establish-
ments and those determined
to be directly connected to a
public supply.