EasyManua.ls Logo

Dynisco 1392 Series - Instrument Configuration

Dynisco 1392 Series
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
1.4 Instrument Configuration
Remove the instrument from case.
Press and hold FUNC button while re-inserting instrument into case.
Continue to hold FUNC button.The display will show 1392, then the Software Version, then
after 2 seconds, the display will show COnF.
Press the FUNC button once. The display will show iP J
Use the UP/DOWN arrows to select the appropriate input type (eg. 4-20mA, 0-10Vdc, etc).
Note: If 0-20mA or 4-20mA is selected, be sure to install the supplied 2.49 ohm resistor
across the V+ and V- terminals.
Press the FUNC button to store the Input type setting and advance to the next parameter.
Using the UP/DOWN arrows set the appropriate display units.
Press FUNC to store the Display Units setting and advance to the next parameter.
Using the UP/DOWN arrows set the appropriate Decimal Point position.
Press FUNC to store the Decimal Point setting and advance to the next parameter.
Display will alternate between F.S.U. and 10000, use the UP/DOWN arrows to set the
appropriate Full Scale Units as specified on the Sensor.
Press FUNC to store the FSU setting and advance to the next parameter.
If further settings need to be modified, refer to Chapter 7 for a full description.
Hold the FUNC button until display returns to Normal Operation.
To set Alarm 1 and 2 thresholds, press the FUNC button until 1.Har is displayed.
Modify the Alarm 1 threshold by using the UP/DOWN arrows until desired value is reached
then press FUNC to store and proceed to Alarm 2 threshold.
Repeat for Alarm 2 threshold and press FUNC to store when done.
Secure the instrument in the case by pushing firmly in the bezel until it clicks into the
retaining tabs.
The Indicator is now ready for use.